June 25, 2021, Filed Under: MEMS processingOxford RIE 80 1.login user: userpassword: user 2.pre-cleaning(If the chamber is already evacuated, we can directly run “O2 clean” recipe; If not, we neet to evacuate the chamber first.) “process”“recipes”“load”“Oxygen Plasma”->edit the processing time to 30 mins or 60 mins.“run” 3.sample loading place your sample on the wipe When the O2 clean is done, blue alert pops up. ->”continue” wait until the “vent time left” less than 600 secs.turn “chamber up” and press two “chamber hoist” simutiniously to open the chamber. load your sample onto the chucktaping the flat edge of a wafer to avoid scratch due to wafer tweezer if you have a film.turn “chamber down” and press two “chamber hoist” buttons to close the chamber. 4.run etching recipe “stop”“evacuate”wait penning pressure less than e-5 Torr.“process”“recipes”“load”select your recipe and “run”when the recipe is done, the chamber is vented automatically. 5.unload sample turn “chamber up” and press two “chamber hoist” buttons to open the chamber.remove your wafer and place it onto the wiper in the glove box.turn “chamber down” and press two “chamber hoist” buttons to close the chamber.“stop”“evacuate”select “cancel” if you don’t load a wafer. you can leave now 🙂