July 12, 2021, Filed Under: MEMS processingOxford ICP RIE Etcher 1.Login user: guestpassword: guest 2.Venting “stop”“vent”. Wait until time is less than 90 secs. 3.dummy wafer loading open the Load Lock and load the dummy wafer.don’t touch the wafer belt.“stop”“evacuate”Pirani guage will be decreased.Penning guage should be decreased until e-5 4.pre-cleaning step “process”“recipes”“load”“O2 chamber clean”left click “process step” and edit the “process time” to “60 min” or “30 min”“run”when the process is done, yellow warning message will come up then press “accept” button 5.dummy wafer unloading “stop”“vent”open the Load Lock and unload the dummy wafer. 6.Sample loading Mount your sample on the surface of the dummy wafer using tap or oil.Load the dummy wafer and your sample to the Load Lock again.“stop”“evacuate”. wait the penning guage until it reaches e-5. 7.recipe running “process”“recipes”“load”“run” the plasma is working. When the recipe is done, the yellow alert shows again. ->”accept” 8.unload your sample “stop”“vent”take your sample out of the Load Lock. “stop”“evacurate” now you are done and you can leave.