PXRD Anton Paar

Congratulations we have finally obtained the state-of-the-art PXRD system The Anton Paar XRDynamic500.

All Specifications

X-ray source
Source type Primux 3000
New generation X-ray generator. Up to 3 kW Spellman with Primux tube Cu source
Tunable voltage range / current 20 kV to 60 kV / 2 mA to 50 mA
Goniometer
Coupled Scans Vertical θ/θ geometry Bragg-Brentano / with Parallel Beam and Monochromator options.
Selectable Measurement radius 360 or 400 mm
Wide usable angular range -95° to 162.5° 2θ
High resolution step size 0.0001°
2θ linearity ≤0.01°
Maximum angular speed 15 °/sec
Maximum angular resolution 0.021° (FWHM of 1st LaB₆ peak)
Sample stages and attachments
Ambient sample stages Fixed sample stage
Sample spinner stage (reflection/transmission)
XY stage (with autosampler option), 12-autosampler
Capillary spinner stage
Non-ambient attachments TTK 600 from -196 to 600 C, innert atmosphere.
Detectors
  • Pixos 2000 detector (0D and 1D mode) with Fluorescence energy discrimination option and ROI
Software
  • XRDdrive: system control and data acquisition software
  • XRDanalysis: data processing and analysis software for qualitative and quantitative phase analysis, microstructure analysis, and Rietveld refinement
  • XRDview: data plotting and visualization package